Aalborg GFC Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Relé Aalborg GFC. Aalborg GFC User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 40
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
Technical Data Sheet No. TD9709M Rev. N
Date of Issue: March 2013
OPERATING MANUAL
GFC MASS FLOW
CONTROLLER
Zobrazit stránku 0
1 2 3 4 5 6 ... 39 40

Shrnutí obsahu

Strany 1 - OPERATING MANUAL

Technical Data Sheet No. TD9709M Rev. NDate of Issue: March 2013OPERATING MANUALGFC MASS FLOW CONTROLLER

Strany 2

64. SPECIFICATIONSFLOW MEDIUM: Please note that GFC Mass Flow Controllers are designed to work withclean gases only. Never try to meter or control f

Strany 3 - TABLE OF CONTENTS

LEAK INTEGRITY: 1 x 10-7sccs He max to the outside environment.ATTITUDE SENSITIVITY: No greater than ±15 degree rotation from horizontal. Standardcali

Strany 4

8TABLE I GFC 17 LOW FLOW MASS FLOW CONTROLLER*4.2 Flow CapacitiesMODEL CODEmL/min [N2]CODEliters/min [N2]1701 10 07 102 20 08 203 50 09 504 100 10 100

Strany 5

GFC 17/37/47 CAUTIONCAUTION: If the valve is left in the AUTO (control) or OPEN (PURGE)mode for an extended period of time, it may become warm or even

Strany 6

5.2 Flow Signal Output ReadingsThe flow signal output can be viewed on the LCD display, remote panel meter, dig-ital multimeter, or other display de

Strany 7 - PIN FUNCTION

Setpoints are controlled locally or remotely. Setpoints inputs respond to analog 0to 5 VDC or 4 to 20 mA reference voltages (default jumper setting is

Strany 8

12,Conversely, when the connection is left open or pin [12] remains unconnected thevalve remains active. The valve will remain active when the VALVE

Strany 9

6.2 Flow Path CleaningInspect visually the flow paths at the inlet and outlet ends of the meter for anydebris that may be clogging the flow through t

Strany 10

14FIGURE 6-1 SOLENOID VALVEVarious corrosive gases may demand more frequent replacement of VITON7O-rings and seals inside the valve. Be sure to use an

Strany 11

7. CALIBRATION PROCEDURESNOTE: Removal of the factory installed calibration seals and/or any adjustments made to the meter, as described in this sec

Strany 13

16FIGURE 7-1 CALIBRATION POTENTIOMETER AND JUMPER LOCATIONS (BACK OF GFC)7.2 Calibration of GFC Mass Flow ControllersAll adjustments in this section

Strany 14

177.2.2 ZERO AdjustmentShut off the flow of gas into the Mass Flow Controller. To ensure that no seepage orleak occurs into the meter, temporarily

Strany 15

7.2.5 Connections and Initial Warm UpConnect the multimeter to output pins [1] and [2] for 0 to 5 VDC (or pins [9] and[14] for 4 to 20 mA) of the 15

Strany 16

7.2.9 75% Flow AdjustmentUsing the flow regulator, increase the flow rate to 75% of full scale flow. Check theflow rate indicated against the flow c

Strany 17

7.2.14 50% Flow AdjustmentChange the setpoint to 2.50 VDC to control at 50% of full scale flow. Check theflow rate indicated against the flow calibr

Strany 18

7.2.17.2 Adjust ScalingUsing a digital multimeter connected to either the 0 to 5 VDC or 4 to 20 mA sig-nal at the 15-pin “D” connector, set the flow r

Strany 19

228.2 Troubleshooting Guide INDICATION LIKELY REASON REMEDYlack of reading power supply off check connection of power supply or outputfuse blo

Strany 20

INDICATION LIKELY REASON REMEDYcalibration off gas metered is not the same as use matched calibrationwhat meter was calibrated forcomposition of gas

Strany 21

24QO2 = Qa = Qr X K = 1000 X 0.9926 = 992.6 sccmwhere K = relative K factor to reference gas (oxygen to nitrogen)1d X Cpwhere d = gas dens

Strany 22

25aalborg7APPENDIX 1COMPONENTS DIAGRAMGFC METERING PC BOARD (TOP SIDE)

Strany 23

TABLE OF CONTENTS1. UNPACKING THE GFC MASS FLOW CONTROLLER...1.1 Inspect Package for External Damage...

Strany 24

26COMPONENTS DIAGRAMMETERING PC BOARD (BOTTOM SIDE)

Strany 25

27APPENDIX 2GAS FACTOR TABLE (“K” FACTORS)CAUTION: K-Factors at best are only an approximation. K factors should not be used in applications that r

Strany 26

28ACTUAL GASK FACTORRelative to N2Cp[Cal/g]Density[g/I]Deuterium D21.00 1.722 1.799Diborane B2H6.4357 .508 1.235Dibromodifluoromethane CBr2F2.1947 .15

Strany 27

29ACTUAL GASK FACTORRelative to N2Cp[Cal/g]Density[g/I]Hydrogen Bromide HBr1.000 .0861 3.610Hydrogen Chloride HCl1.000 .1912 1.627Hydrogen Cyanide HCN

Strany 28

30ACTUAL GASK FACTORRelative to N2Cp[Cal/g]Density[g/I]Phosphorous Oxychloride POCl3.36 .1324 6.843Phosphorous Pentafluoride PH5.3021 .1610 5.620Phosp

Strany 29 - COMPONENTS DIAGRAM

31APPENDIX 3DIMENSIONAL DRAWINGS2.380.95DIMENSIONS: INCH [mm]GFC17 MASS FLOW CONTROLLERNOTE: Aalborg7 reserves the right to change designs and dimensi

Strany 30

322.69 [68.3]0.69 [17.5]3.94 [99.9]4.88 [123.8]5.98 [151.8]FOR HIGH FLOW MASS FLOW CONTROLLER ONLY6-32DIMENSIONS: INCH [mm]2.380.951.25 [31.8]0.28 [7.

Strany 31

330.95 SAE/MS SWAGELOK 3/8 TUBE CONNECTOR 2.38 7" (177,8) 9.98 (253,5) 12.30 (312,4) 10-24 UNC 0.250.18 (4,6)1.39 (35,3)1.39 (35,3)1" (25

Strany 32

347.55 (191.8 mm) 3.00 (76.2 mm) 3.00(76.2 mm)12.62 (320.5 mm)10.42 (264.7 mm)7.25 (184.1 mm)2.50 (63.5 mm)6.75 (171.5 mm)4 x 1/4-20 UNC-2B 0.352 x

Strany 33

35NOTE: Aalborg7 reserves the right to change designs and dimensions at its sole discretion at any time without notice. For certified dimensions pleas

Strany 34

7.2.5 Connections and Initial Warm Up...7.2.6 ZERO Adjustment...

Strany 35 - GFC17 MASS FLOW CONTROLLER

APPENDIX 4NOTE: Follow Return Procedures In Section 1.3.WARRANTYAalborg7 Mass Flow Systems are warranted against parts and workmanship for a period o

Strany 36 - GFC37/47 MASS FLOW CONTROLLER

11. UNPACKING THE GFC MASS FLOW CONTROLLER1.1 Inspect Package for External DamageYour GFC Mass Flow Controller was carefully packed in a sturdy card

Strany 37 - 1.39 (35,3)

2. INSTALLATION2.1 Primary Gas ConnectionsPlease note that the GFC Mass Flow Controller will not operate with liquids. Onlyclean gases are allowed

Strany 38 - GFC67 MASS FLOW CONTROLLER

3PIN FUNCTION1 0 to 5 VDC Flow Signal Common2 0 to 5 VDC Flow Signal Output3 Common4 Open (Purge)5 Common, Power Supply6 (unassigned)7 +12 VDC (+24 VD

Strany 39 - GFC77 MASS FLOW CONTROLLER

FIGURE 2-2, POTENTIOMETER AND JUMPER LOCATIONS2.2.1 Valve Control ConfigurationThere are three basic valve control options. (a) LOCAL or REMOTE contr

Strany 40 -  CAUTION:

2.2.3 Panel Mounting ReadoutsAnother option for the GFC Mass Flow Controller is the Panel Mounting RemoteReadout. In this configuration the LCD reado

Komentáře k této Příručce

Žádné komentáře