d) Parameters configuration
The following parameters must be configured:
Device Function (See 5.3.12 Submenu “Device Function”).
“Controller”
function must be selected.
Full Scale Range (See 5.3.14 Submenu “Device Calibration”). Full
Scale Range parameter must be set equal to the GFC full scale flow
rate in Litr/min.
Fluid Std. Density (See 5.3.14 Submenu “Device Calibration”). This
parameter is required only when mass-based engineering units are
selected.
User may configure other parameters (see Paragraph 5.3) according to his pref-
erences and application requirements.
6.2.3 Connecting TIO to flow meters/controllers
from other manufactures (stand alone)
a) Mounting
When TIO is connected to flow meters / controllers from other
manufactures, it can be used as stand-alone table top or panel-
mounted (see Figure 6.8). On the back side of the TIO enclosure there
are 4 tapped holes which are designated to be used for the panel-
mounted option.
NOTE: If “Full Scale Range”, “Device Function” and “Fluid Std.
Density” parameters are not set properly, the device may have
erroneous reading and unpredictable behavior.
,