78
AALBORG7 TOTALIZER-IO ASCII SOFTWARE INTERFACE COMMANDS
NOTE: AN "*" INDICATES POWER UP DEFAULT SETTINGS.
A
N "**"INDICATES OPTIONAL FEATURE NOT AVAILABLE ON ALL MODELS.
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND ARGUMENT 1 ARGUMENT 2 ARGUMENT 3 ARGUMENT 4 RESPONSE
Flow Requests the current flow
reading in current EU
1
F
<Value> (Actual flow in
current engineering units)
Set Point
Set point value in current EU
(Read, Write) With Argument
#2 = ‘S’ set point value will be
saved in the EEPROM
2 S
NO ARGUMENT
(read status)
S:<Value> (Set Point in
current engineering units)
Example: S:20.5
<Value> (write,
not saved in EEP-
ROM memory)
S:<Value> (Set Point in
current engineering units)
Example: S:20.5
<Value>(write,
saved in EEPROM
memory)
S S:<Value> ,S (Set Point in
current engineering units)
Example: S:20.5,S
Density
Read / Set Fluid Density for
standard conditions in g/litr
[0.000001-10000.0] g/litr
3 D
NO ARGUMENT
(read current
value)
D:<Value> (Actual density
in g/litr)
Example: D:1.25
<Value>
(write and save
new value)
D:<Value> (Actual density
in g/litr)
Example: D:1.56
Diagnostic
Events
Register
Read/Reset current status of
Diagnostic Events Register
See list of the Diagnostic
Events below.
4
DE
NO ARGUMENT
(read status)
DE:0x10 0x10 – diagnostic
word (16 bits wide)
R (reset Event
Log register to
0x0000)
DE:0x0